Atomic Force Microscopy
A sharp tip on a springy beam, a laser to read how the beam bends, and a feedback loop that keeps the force constant. Everything else in AFM follows from those three parts and the physics of the last nanometre.
Optical-lever detection and the z-feedback loop
The tip sits at the free end of a micro-fabricated cantilever. A laser reflects off the back of the lever onto a four-quadrant photodiode (PSD). Vertical bending moves the spot up or down (the A+B − (C+D) signal); torsion moves it sideways (the (A+C) − (B+D) signal, used in lateral force microscopy). A controller drives the z-piezo to keep the chosen signal at its setpoint, and the z-voltage it needs at every pixel is the height image.
The lever is its own lens
A tip deflection Δz tilts the free end by 3Δz/2L. The reflected beam turns by twice that, so at distance D the spot moves 3DΔz/L. With D ≈ 3 cm and L ≈ 100 µm, that is a gain of about 900.
Piezo scanner
Tube or flexure stages move by nanometres per volt. Open-loop piezos creep and show hysteresis; closed-loop scanners with capacitive or strain sensors trade a little noise for linear, repeatable motion.
Three channels at once
Height (z-voltage), error (deflection or amplitude residual, a map of edges) and phase in dynamic modes. Always save trace and retrace; their disagreement tells you how well the loop tracked.
A spring, a resonator, a thermometer
For a rectangular beam clamped at one end, Euler–Bernoulli theory gives the spring constant and the fundamental flexural resonance from geometry and material alone. Stiffness scales with \(t^3/L^3\), so thickness is the parameter that manufacturers control worst and that matters most: a 10 % thickness error is a 33 % error in \(k\). Always calibrate (§7) rather than trusting the nominal value.
Cantilever calculator
What the tip actually feels
Long-range forces (van der Waals, electrostatic, magnetic, capillary) pull the tip in over nanometres. Short-range chemical and Pauli repulsion forces act over ångströms and carry atomic contrast. When the tip touches, continuum contact mechanics takes over: Hertz for elastic contact, and DMT or JKR once adhesion matters. Which adhesion model applies is set by the Tabor parameter \(\mu\): small, stiff, weakly adhering contacts are DMT-like (\(\mu \lesssim 0.1\)); large, soft, sticky ones are JKR-like (\(\mu \gtrsim 5\)).
Force–distance simulator
DMT tip–sample law · quasi-static leverLeft: the tip–sample force. Where its slope exceeds the lever stiffness (shaded), the tip cannot sit still and snaps in. Right: what you record while ramping the piezo, the classic approach and retract curve. The hysteresis loop area is the work lost per cycle; the retract minimum is the pull-off force.
Four ways to hold the tip near a surface
Modes differ in what the feedback loop regulates. In static mode it is the deflection itself. In dynamic modes the lever oscillates and the loop regulates amplitude, frequency shift or the peak force of each tap. The choice sets the force you apply, the speed you can reach and what else you can measure.
- Regulates
- Static deflection, i.e. a constant normal force
- Regime
- Continuous repulsive contact
- Typical lever
- k 0.01–1 N/m · f₀ 5–50 kHz
- Best for
- Flat, hard samples; lateral force (friction), conductive AFM, piezoresponse, lithography
- Watch for
- Lateral shear drags soft samples and blunts tips; 1/f noise and drift sit right in the signal band
- Regulates
- Oscillation amplitude A at a fixed drive near f₀; setpoint A/A₀ ≈ 0.6–0.9
- Regime
- Intermittent contact; attractive or repulsive depending on A₀ and setpoint
- Typical lever
- k 10–50 N/m · f₀ 200–400 kHz · A 5–50 nm
- Best for
- General imaging in air and liquid; phase maps material contrast via dissipation
- Watch for
- Bistability between regimes (height and phase jumps); slow response when Q is high (vacuum)
- Regulates
- Frequency shift Δf. A phase-locked loop tracks the resonance; an amplitude loop keeps A constant, and its drive signal measures dissipation
- Regime
- Mostly attractive; short-range forces at the lower turning point
- Typical lever
- Si: k 30–50 N/m, A 5–20 nm · qPlus: k ≈ 1800 N/m, A < 100 pm
- Best for
- True atomic and sub-molecular resolution in UHV; quantitative force spectroscopy (Sader–Jarvis inversion)
- Watch for
- Needs vibration, thermal and electronic stability; Δf is a non-monotonic function of distance, so feedback can crash on the wrong branch
- Regulates
- Peak force of each tap. The z-piezo is modulated far below resonance (≈0.1–10 kHz) and a full force curve is recorded at every pixel
- Regime
- Intermittent contact with direct force control, often tens to hundreds of pN
- Typical lever
- k 0.1–5 N/m (up to ~200 N/m for stiff samples)
- Best for
- Soft and biological samples; maps of modulus, adhesion, deformation and dissipation (e.g. PeakForce Tapping, fast force–volume)
- Watch for
- Modulus values are only as good as the contact model, tip radius and k calibration you feed them
The cantilever as a driven, damped oscillator
Near its first resonance the lever behaves like a point-mass oscillator. A tip–sample force gradient acts as an extra spring: attraction softens the system and lowers the resonance, repulsion stiffens it. In AM-AFM you drive at fixed frequency and read the resulting amplitude drop; in FM-AFM you track the resonance and read its shift directly. The small-amplitude result is the one to remember; for amplitudes larger than the force range, use the Giessibl integral.
Resonance explorer
drive fixed at the free resonance, as in AM-AFMEvery image is the sample dilated by the tip
The recorded height at each x is where the tip apex ends up when the tip body first touches the surface. Mathematically that is a grey-scale dilation of the true surface by the reflected tip shape. It widens protrusions, narrows and shallows holes and gaps, and cannot be fully undone: regions the tip never touched are lost. Heights of isolated objects on a flat substrate survive; lateral sizes do not.
Tip convolution sandbox
spherical apex on a cone · vertical scale exaggeratedFrom approach to a publishable image
- Choose the leverMatch k and f₀ to mode and sample (§2). Soft sample, soft lever.
- Align the laserMaximise the PSD sum on the lever end; centre the spot; zero deflection.
- CalibrateThermal tune for f₀, Q and k; InvOLS on a hard reference. Log both.
- Approach gentlyConservative setpoint (A/A₀ ≈ 0.9 or low peak force), small scan size first.
- Tune the loopRaise I then P until trace and retrace overlap without ringing.
- Pick pixels and ratePixel size at most half the smallest feature; tip speed = 2 × size × line rate.
- Process lightlyPlane or line flatten with features masked; avoid filters that invent structure.
- ReportInstrument, lever type, k, f₀, mode, setpoint, scan rate, pixels, processing steps.
Turning volts into newtons
Deflection sensitivity. Push the tip onto a surface much stiffer than the lever (sapphire, silicon, clean glass). In contact the lever deflects exactly as much as the piezo moves, so the slope of the PSD voltage against z gives the InvOLS in nm/V. Re-measure after every laser realignment and in every medium.
Spring constant. The thermal method fits the first thermal resonance peak and applies equipartition, with a correction because the optical lever measures slope, not displacement, and was calibrated with an end-loaded static bend. The Sader method uses the plan-view geometry, resonance and Q in a fluid of known density and viscosity, and does not touch the surface.
The thermal limit. Brownian motion of the lever sets a floor on the smallest force and force gradient you can measure. Both improve with softer levers, higher Q and narrower bandwidth. High Q, however, slows AM-AFM (§5), which is why FM detection is preferred in vacuum.
Typical uncertainty. Careful thermal calibration gives k to about 5–10 %; nominal datasheet values can be off by a factor of two or more.
The tip as a local probe of other properties
Kelvin probe force microscopy
An ac bias makes the electrostatic force oscillate at ω with amplitude proportional to the potential difference. A dc bias nulls it, and that dc value maps the contact potential difference (work function).
Magnetic force microscopy
A magnetised tip is lifted 20–100 nm on a second pass. The phase or frequency shift follows the stray-field gradient, which reveals domains and vortices.
Conductive AFM
A biased conductive tip in contact maps current at fixed load. Spreading resistance and the tunnel barrier dominate, so contact force and tip coating must be kept constant.
Piezoresponse force microscopy
An ac bias through the tip makes a piezoelectric or ferroelectric surface expand and contract. Amplitude gives the local coefficient; phase gives polarisation direction.
Lateral and friction force
Lever torsion during a scan perpendicular to its axis measures friction. Half the trace–retrace difference cancels topographic crosstalk.
Single-molecule force spectroscopy
Pull a tethered molecule and fit the extension with the worm-like chain. Rupture forces grow with the logarithm of the loading rate r, as the Bell–Evans model predicts.
Nanomechanical mapping
Fit each force curve with a contact model to map Young's modulus. Use Hertz for a sphere at shallow indentation and Sneddon for a cone at deep indentation. Watch for substrate effects when δ is more than about 10 % of the film thickness.
Bond-resolved imaging
A stiff quartz tuning-fork sensor oscillates with sub-ångström amplitude, and a CO-functionalised tip resolves the chemical structure of single molecules. Compare data across sensors using the normalised frequency shift.
High-speed AFM
Small levers with high f₀ and low k, fast scanners and dynamic PID reach video rate (>10 frames/s) on proteins in liquid. The loop bandwidth sets the ceiling.
Diagnose the image before you believe it
| What you see | Likely cause | What to do |
|---|---|---|
| Features wider than expected, holes shallow, gaps closed affects every image | Tip convolution; blunt or worn apex | Sharper tip; image a tip characteriser; report heights, not widths; blind tip reconstruction |
| Every feature repeated at the same offset discard data | Double or multiple tip | Replace the tip; an offset that rotates with scan angle confirms it |
| Ringing or overshoot at step edges | P or I gain too high; loop oscillates | Lower gains until the error signal is quiet on flat areas |
| Trace and retrace disagree; tails on the downhill side | Gains too low or scan too fast; tip loses contact ("parachuting") | Raise I gain, slow the line rate, lower the setpoint slightly |
| Sudden streaks along the fast axis | Debris picked up or dropped by the tip; tip instability | Rescan; lower force; clean sample; change tip if persistent |
| Bowl or saddle over the whole frame | Scanner bow (tube pendulum motion); sample tilt | Polynomial flattening with features masked; closed-loop flexure scanner |
| Compressed or stretched start of each frame worst after large offsets | Piezo creep and hysteresis | Let the scanner settle; overscan; closed-loop mode; calibrate on a grating |
| Features slowly slide between frames | Thermal drift | Equilibrate 30–60 min after laser on and loading; drift-correct with fiducials |
| Fine periodic fringes on reflective samples | Optical interference of stray laser light | Realign the spot on the lever end; use dynamic mode; tilt the sample |
| Height or phase contrast inverts mid-image tapping mode | Switching between attractive and repulsive regimes | Change A₀ or setpoint to stay in one regime; check the phase: above 90° attractive, below 90° repulsive |
| Soft objects too low; height depends on setpoint | Sample compressed by excess force | Reduce force; off-resonance tapping with peak force control; measure height vs. force |
| Lines at 50/60 Hz or at a fixed frequency | Electrical pickup, acoustic or building vibration | Check grounding, acoustic hood, active isolation; compare the PSD noise spectrum |
From thermal kicks to breaking bonds
Order-of-magnitude forces met in AFM experiments, on a logarithmic axis. Molecular unbinding forces depend on loading rate, so treat those as representative values.
Forty years in ten entries
- 1982Scanning tunnelling microscope: the first real-space images of surface atoms. Binnig, Rohrer, Gerber, Weibel · Nobel Prize 1986
- 1986Atomic force microscope proposed and demonstrated, with an STM reading the lever deflection. Binnig, Quate, Gerber · PRL 56, 930
- 1988Optical beam deflection detection makes AFM practical and commercial. Meyer & Amer · APL 53, 1045
- 1991Kelvin probe force microscopy and FM detection for non-contact AFM. Nonnenmacher et al. · Albrecht et al.
- 1993Tapping mode brings gentle, reliable imaging in air and soon in liquid. Zhong, Inniss, Kjoller, Elings
- 1993Thermal-noise spring-constant calibration. Hutter & Bechhoefer · RSI 64, 1868
- 1995True atomic resolution of Si(111)-7×7 by non-contact AFM. Giessibl · Science 267, 68
- 1997Protein unfolding by single-molecule force spectroscopy (titin). Rief et al. · Science 276, 1109
- 2001High-speed AFM for biomolecules at video rate. Ando et al. · PNAS 98, 12468
- 2009Chemical structure of pentacene resolved with a CO-terminated tip. Gross, Mohn, Moll, Liljeroth, Meyer · Science 325, 1110
Where to go next
- G. Binnig, C. F. Quate, Ch. Gerber, "Atomic force microscope," Phys. Rev. Lett. 56, 930 (1986). doi
- F. J. Giessibl, "Advances in atomic force microscopy," Rev. Mod. Phys. 75, 949 (2003). doi
- R. García, R. Pérez, "Dynamic atomic force microscopy methods," Surf. Sci. Rep. 47, 197 (2002). doi
- H.-J. Butt, B. Cappella, M. Kappl, "Force measurements with the atomic force microscope: Technique, interpretation and applications," Surf. Sci. Rep. 59, 1 (2005). doi
- T. R. Albrecht, P. Grütter, D. Horne, D. Rugar, "Frequency modulation detection using high-Q cantilevers for enhanced force microscope sensitivity," J. Appl. Phys. 69, 668 (1991). doi
- J. L. Hutter, J. Bechhoefer, "Calibration of atomic-force microscope tips," Rev. Sci. Instrum. 64, 1868 (1993). doi
- H.-J. Butt, M. Jaschke, "Calculation of thermal noise in atomic force microscopy," Nanotechnology 6, 1 (1995).
- J. E. Sader, J. W. M. Chon, P. Mulvaney, "Calibration of rectangular atomic force microscope cantilevers," Rev. Sci. Instrum. 70, 3967 (1999). doi
- J. E. Sader, S. P. Jarvis, "Accurate formulas for interaction force and energy in frequency modulation force spectroscopy," Appl. Phys. Lett. 84, 1801 (2004). doi
- J. P. Cleveland, B. Anczykowski, A. E. Schmid, V. B. Elings, "Energy dissipation in tapping-mode atomic force microscopy," Appl. Phys. Lett. 72, 2613 (1998).
- J. S. Villarrubia, "Algorithms for scanned probe microscope image simulation, surface reconstruction, and tip estimation," J. Res. NIST 102, 425 (1997).
- M. Nonnenmacher, M. P. O'Boyle, H. K. Wickramasinghe, "Kelvin probe force microscopy," Appl. Phys. Lett. 58, 2921 (1991). doi
- L. Gross et al., "The chemical structure of a molecule resolved by atomic force microscopy," Science 325, 1110 (2009). doi
- T. Ando et al., "A high-speed atomic force microscope for studying biological macromolecules," PNAS 98, 12468 (2001). doi
- Textbooks: R. García, Amplitude Modulation Atomic Force Microscopy (Wiley-VCH, 2010); P. Eaton, P. West, Atomic Force Microscopy (Oxford UP, 2010); S. Morita et al. (eds.), Noncontact Atomic Force Microscopy, vols. 1–3 (Springer).
Simulations use idealised models (point-mass oscillator, DMT contact, rigid spherical-conical tip) meant for intuition, not quantitative fitting. Material constants are representative room-temperature values.